LFY-401D fabric induction electrostatic tester

LFY-401D fabric induction electrostatic tester

The main purpose
Under laboratory conditions, this instrument is suitable for measuring the electrostatic properties of fibers, yarns, fabrics, carpets, decorative fabrics and other types of fabrics or various plate-shaped products. The instrument host is composed of a corona discharge device and a probe detector. Use a given high-voltage electric field to discharge the fabric for a fixed time to induce static electricity on the fabric, so as to detect the amount of static electricity, the half-life of static voltage attenuation, and the amount of static electricity remaining to show the electrostatic performance of the tested fabric. The instrument is equipped with a constant temperature and humidity box to provide suitable temperature and humidity conditions for the electrostatic test.
Instrument characteristics
1. Whether the fabric is seriously charged or not depends on the level of the static voltage after the high-voltage discharge on the one hand, and on the other hand the static voltage holding time. The electrostatic escaping speed on the fabric is expressed by the electrostatic half-life, which is an important assessment index for testing the electrostatic performance of the fabric.
2. The instrument is composed of a host computer and a computer detection system, which can be tested by the timing method and the constant pressure method respectively.
3. The instrument uses a computer system to directly display the instantaneous peak value and half-life value of the fabric during high-voltage discharge;
4. Real-time display of the voltage curve with time.
5. Automatic shutdown of high voltage and motor, easy operation.
6. The instrument has heating, cooling, humidifying, dehumidifying and timing functions.
7. Temperature and humidity can be set on the instrument panel as required.
Technical index
1. The measuring range of electrostatic voltage value is 0~10kV
2. The half-life time limit range is 0~99.99h
3. Turntable speed 1500 rpm
4. Discharge time: 0-99.99s (the test is set at (30±0.1) seconds)
5. Discharge distance between needle electrode and sample: 20mm
6. The measuring distance between the test probe and the sample: 15mm
7. Temperature adjustment range: 20℃~50℃, accuracy: ±1.5℃, display accuracy: 0.1℃
8. Relative humidity adjustment range: 25%RH~90%RH, accuracy ±3%RH, display accuracy: 0.1℃
9. Timing range: 0-24h, accuracy: ±5s
10. Effective sample size: 45mm×45mm
11. The instrument has been measured and corrected for static voltage before it is sent out
12. Check as required before use, the oil cylinder in the main engine must be full of oil (transformer oil)
13. Power supply: 220V 50Hz 1.5kW
Applicable standards (for different standards, different accessories are required)
GB/T 12703.1-2008 Textiles-Evaluation of electrostatic properties-Part 1: Static voltage half-life
FZ/T 01042-1996 Textile materials Electrostatic properties Determination of electrostatic voltage half-life
GB/T 33730-2017 Evaluation of electrostatic properties of hair products
Baidu