The main purpose
This instrument is used to detect the slippage and loss of non-woven materials for specific test fluids.
Instrument characteristics
1. It has a drain station and sample fixture that meet the requirements of the standard;
2. It has a dosing device that can give 25mL test liquid in a continuous flow within (4±0.1)s;
3. Manual operation, automatic recording and digital display of the outflow time of the test liquid;
4. It can be equipped with a loss weighing system (need to be purchased separately).
Technical index
1. Flow rate control range: 6.10mL/s~6.41 mL/s.
2. Sliding and drain angle control: 25°.
3. Length of slip and drain: (250±0.2) mm.
4. Timing: 60s, accuracy: ±0.1s.
Applicable standards
YY/T 0506.3-2005 Surgical drapes, surgical gowns and clean clothes for patients, medical staff and instruments-Part 3: Test methods
GB/T 24218.11 Textiles Nonwovens Test Methods Part 11: Measurement of Overflow
Edana152.0-99
ISO 9073-11