Hello! Welcome to the website of Shandong Textile Research Institute!

24-Hour Service Hotline

13475428416  15954827913

Advanced Search:
Latest Announcement

Test methods for primary wound dressings - Waterproofness

Source: Time:2021-05-08 08:41:00 views:

Test conditions
The test sample should be adjusted for at least 16 hours under the condition of (21±2)℃ and relative humidity (RH) of (60±15)%. And test under the same conditions.
instrument
1 Water resistance measuring instrument: There is a pool that can apply 500mm hydrostatic pressure for 300s to the circular area of the non-adhesive surface of the test material. The sample is fixed in a horizontal position with two rings, and the lower ring is an integral part of the tank. The screw device is used for clamping to prevent leakage and sample movement during the test. The hydrostatic pressure is generated by a vertical pipe connected to the bottom of the pool with an inner diameter of not less than 3mm.
To prevent leakage, the annular surface in contact with the sample should be covered with a suitable covering material, such as rubber.
2 Distilled water, deionized water.
3 steps
3.1 The prepared sample should be free of wrinkles.
3.2 Fill the pond with purified water at (21±2)℃.
3.3 Place the sample on the lower ring in a horizontal sliding manner to avoid air between the water surface and the lower surface of the sample.
3.4 Cover the upper surface of the sample with dry filter paper larger than the test area, put on the upper ring, and clamp it with a screw device.
3.5 Fill the tube with water to reach the required water level above the sample surface.
3.6 Maintain the hydrostatic pressure (300 ± 10) s.
3.7 Check whether there is water seeping through the sample on the filter paper and record the result.

The LFY-244D medical material Waterproofness tester produced by Shandong Textile Research Institute meets the standard YY/T 0471.3-2004 Test methods for primary wound dressings Patrt 3 - Waterproofness

LFY-244D医用材料阻水性能测试仪

The LFY-244D medical material Waterproofness tester 

Home  |  About  |  News  |  Products  |  Honor  |  Testing Standard  |  Special education  |  New Products  |  Recruitment

 Address: 195 Shandong Road, Qingdao   Postcode: 266032
Tel: 0532-85614755 (Sales switchboard) 0532-85635588  (sales manager) 
Fax: 0532-85653787  He Hongxia (Senior Engineer) Deputy Director   15954827913  QQ: 179584685
After-sales service tel: 0532-85624778  15192792891

鲁ICP备10203291号-10       鲁公网安备 37020302370559号     Support: LEIXUN

Baidu
map