Hello! Welcome to the website of Shandong Textile Research Institute!

24-Hour Service Hotline

13475428416  15954827913

Advanced Search:
Medical Testing Equipment
LFY-718Protective clothing anti synthetic blood and bacteriophage penetration tester

Name:LFY-718Protective clothing anti synthetic blood and bacteriophage penetration tester

Details

The main purpose

The protective clothing anti-virus penetration tester is suitable for testing the penetration ability of protective clothing materials against blood-borne pathogens and the ability of protective materials to resist liquid penetration.

working principle

A suspension containing substitute microorganisms was used to measure the virus penetration ability under a certain hydrostatic pressure.

Technical index

1. The instrument uses an air source that can provide (20-22) kPa air pressure to continuously pressurize the sample without being restricted by the space of the test site.

2. The instrument has a pressure gauge to display the pressure, which is adjustable.

3. Use pressurized medium: compressed air.

4. The special stainless steel penetrating test tank ensures that the sample is firmly clamped and prevents the test liquid from splashing around.

5. Transparent cover material: corrosion resistance.

6. Square metal blocking net: open space ≥50%; deformation ≤5mm.

7. Pressurization time can be set, digital display timer, accuracy ±1 second.

8. The instrument has a clamp that can produce 13.6N·m torque.

9. The structure of the instrument is the same as the "Schematic diagram of the test instrument" and "The structure of the test tank" in the standard.

10. Pressure point: 1.75 kPa; 3.5kPa; 7kPa; 14kPa; 20kPa

11. Prepared by the user: culture medium, control, phage suspension, sedimentation plate.

12. User-supplied: Thickness gauge, incubator, water bath, balance, vortex gas mixture, refrigerator, sterilizer, oscillator, PH meter, inoculation loop, spectrophotometer, centrifuge, pipette, test tube, Glass bottle.

13. Power supply: AC220V, 50Hz, 100W

Applicable standards (for different standards, different accessories are required)

YY/T 0689-2008 Blood and body fluid protective equipment, protective clothing material resistance to blood-borne pathogen penetration test Phi-X174 bacteriophage test method

YY/T 0699-2008 Liquid chemical protective equipment and protective clothing materials anti-pressurized liquid penetration performance test method

YY/T 1799-2020 Technical requirement of re-usable protective clothing for medical use 4.2.2.1

ISO 16604:2004

ASTM F1670
Standard Test Method for Resistance of Materials Used in Protective Clothing to Penetration by Synthetic Blood
ANSI/ASTM F1671
Test Method for Resistance of Protective Clothing Materials to Penetration by Blood-Borne Pathogens Using Viral Penetration as a Test System
YY/T 1871-2023  Isolation gowns for medical use

YY/T1887-2023   Positive pressure protective clothing ( PPPC) for medical use

T/CTES 1013-2019 Technical specification for classification and performance of protective apparel and surgical drape

GB 19082-2023 Disposable protective clothing for medical use

Home  |  About  |  News  |  Products  |  Honor  |  Testing Standard  |  Special education  |  New Products  |  Recruitment

 Address: 195 Shandong Road, Qingdao   Postcode: 266032
Tel: 0532-85614755 (Sales switchboard) 0532-85635588  (sales manager) 
Fax: 0532-85653787  He Hongxia (Senior Engineer) Deputy Director   15954827913  QQ: 179584685
After-sales service tel: 0532-85624778  15192792891

鲁ICP备10203291号-10       鲁公网安备 37020302370559号     Support: LEIXUN

Baidu
map