24-Hour Service Hotline
13475428416 15954827913
The main purpose
It is used to test the resistance to contact heat of welders' gloves that protect hands and wrists during welding and related operations.
Instrument characteristics
1. It has a heating body that can withstand temperatures up to 500℃, and the temperature is controllable;
2. The special calorimeter meets the test accuracy requirements;
3. It can digitally display the heating body temperature, calorimeter temperature, and critical time;
4. The contact speed is adjustable.
Technical parameter
1. The contact surface diameter of the heating body: (25.2±0.05) mm;
2. The diameter of the calorimeter test surface: (25±0.05) mm;
3. Contact force during measurement: (49±0.5)N;
4. Critical time timer: 0-99.99 seconds, accuracy: ±0.01 seconds;
5. Contact speed: (5.0±0.2) mm/s;
6. Sample size: 80mm in diameter.
Applicable standards (for different standards, different accessories are required)
GB 38453-2019 Protective clothing-Protective clothing against heat, 5.19, Appendix C
GB 8965.3-2022 Protective clothing —Clothing to protect against molten metal splash
GB/T 38306-2019 Hand protection-Protective gloves against thermal risks
FZ/T 73040-2010 Protective gloves for working in the hot environment
AQ 6103-2007 Protective gloves for welders Appendix E Contact heat resistance test method
EN 702 Protective clothing-Protection against heat and flame-Test method-Determination of the contact heat transmission through protective clothing or its materials
EN 407 Protective gloves against thermal risks (heat and/or fire)
EN 12477 Protective gloves for welders
Address: 195 Shandong Road, Qingdao Postcode: 266032
Tel: 0532-85614755 (Sales switchboard) 0532-85635588 (sales manager)
Fax: 0532-85653787 He Hongxia (Senior Engineer) Deputy Director 15954827913 QQ: 179584685
After-sales service tel: 0532-85624778 15192792891
鲁ICP备10203291号-10 鲁公网安备 37020302370559号 Support: LEIXUN