24-Hour Service Hotline
13475428416 15954827913
Main uses
Used to test the radiation thermal resistance and radiation thermal performance of single-layer and multi-layer fabrics under continuous and constant radiation heat sources. Quantitatively evaluate the radiation thermal protection performance of clothing materials through a test with controllable radiation heat flow rate and radiation duration.
Technical indicators and performance parameters
1. The basic components of the instrument include: a vertically placed radiation heat source, a sample clamping component, a pneumatic water-cooled plate, a heat sensor, and a data acquisition/burn analysis system.
2. The radiation heat source of the instrument is an infrared quartz lamp, with a luminous length of 127mm and an average total length of 222mm. The input power can be controlled.
3. The voltage meter accuracy of the instrument is ± 1V.
4. The sensor of the instrument is a circular copper plate heat flux meter with a diameter of 4 ± 0.05cm. The temperature measurement range is from 0 ℃ to 1600 ℃, with an accuracy of ± 0.5 ℃.
5. The instrument can provide at least 2 radiation heat exposure conditions, namely 21 kW/m2 and 84 kW/m2, with an accuracy of ± 2%. And can monitor the readings of the voltmeter to control the level of heat exposure during the testing process.
6. The instrument is equipped with a data acquisition/burn analysis system that can record the temperature response of the calorimeter, calculate the accumulated radiation heat exposure value of the sample, and determine the end time of the experiment by comparing the accumulated heat energy with the experimental performance curve over time. The thermal protection characteristics of flame-retardant textile materials exposed to standard radiation heat can be measured relative to the predicted second degree skin burn.
7. The data acquisition component of the instrument has a minimum sampling rate of 4 samples per second at 250 ° C, and a minimum temperature resolution of 0.1 ° C with an accuracy of ± 0.75 ° C. Capable of correcting cold junctions and converting millivolt thermocouple signals into temperature.
8. The instrument uses a 12 bit high-speed A/D to sample the sensor, with high sampling accuracy and fast speed.
9. The instrument has: overvoltage and overcurrent protection, abnormal actuator protection, communication interruption protection, etc.
10. The instrument has a sensor disconnection alarm function.
11. Radiation heat flow measurement accuracy ± 3%.
12. All pipelines inside the test chamber are protected by high-temperature and flame-retardant pipes.
13. The integrated design of the fume hood and instrument body, equipped with exhaust function, can timely discharge the smoke and exhaust gas generated during the testing process.
14. Power supply: 200V/50Hz 20A
Applicable standards
ASTM F193,ASTM F2702-08
Optional accessories
Air compressor
Address: 195 Shandong Road, Qingdao Postcode: 266032
Tel: 0532-85614755 (Sales switchboard) 0532-85635588 (sales manager)
Fax: 0532-85653787 He Hongxia (Senior Engineer) Deputy Director 15954827913 QQ: 179584685
After-sales service tel: 0532-85624778 15192792891
鲁ICP备10203291号-10 鲁公网安备 37020302370559号 Support: LEIXUN